CCA Channel Instructions
Eliezer's channel:
Aharoni channel:
- Make sure both the power supply and anemometer are off
- Make sure the probe is not connected yet
- Plug in power supply
- Move mechanical switch to N (Neutral)
- Turn on power supply and anemometer
- Change the test mode to N/ADJ
- Connect probe to anemometer with the shortage probe on
- Move mechanical switch to adjust
- Turn the null until the digital voltage reads zero
- Move mechanical switch to N (Neutral)
- Replace the shortage probe with the real probe (make sure the probe holder is secured
- Move mechanical switch to adjust
- Use decades to get the digital voltage back to zero. This should be the same resistance you measured of the probe alone.
- Move mechanical switch to N (Neutral)
- Turn damping to get digital voltage back to zero.
- Check which filter you'd like to use
- Move mechanical switch to operate
- Change the test mode to operate
- Chose gain and offset desired (+-10V).
- When done: Move mechanical switch to N (Neutral)