CCA Channel Instructions

Eliezer's channel:

Aharoni channel:

  1. Make sure both the power supply and anemometer are off
  2. Make sure the probe is not connected yet
  3. Plug in power supply
  4. Move mechanical switch to N (Neutral)
  5. Turn on power supply and anemometer
  6. Change the test mode to N/ADJ
  7. Connect probe to anemometer with the shortage probe on
  8. Move mechanical switch to adjust
  9. Turn the null until the digital voltage reads zero
  10. Move mechanical switch to N (Neutral)
  11. Replace the shortage probe with the real probe (make sure the probe holder is secured
  12. Move mechanical switch to adjust
  13. Use decades to get the digital voltage back to zero. This should be the same resistance you measured of the probe alone.
  14. Move mechanical switch to N (Neutral)
  15. Turn damping to get digital voltage back to zero.
  16. Check which filter you'd like to use
  17. Move mechanical switch to operate
  18. Change the test mode to operate
  19. Chose gain and offset desired (+-10V).
  20. When done: Move mechanical switch to N (Neutral)