User Tools

Site Tools


wave_gauges_calibration_procedure

WAVE GAUGES CALIBRATION PROCEDURE

Turning on the equipment:

  • A) Switch on the electrical boxes placed on the stage
  • B) Make sure the Ethernet Hub, NI cDAQ-9184 & Amplifier (Figure ##REF:Fig1##) are powered on.
    When using the old amplifier, please turn it on at least 1 hour before measurement


A) Top - NI cDAQ-9184 ; Bottom - Ethernet Hub ; B) DHI Amplifier
  • C) To check if the DAQ Device is connected correctly:
    open NI MAX (Version 17.5.0f0) ⇒ Devices and interfaces ⇒ Network Devices ⇒ right click on cDAQ9184-1B560CB and select the “self-test” option
    - If completed successfully⇒ click on cards NI 9212, NI 9401, NI 9205 to enable them.
    - If not successful⇒ right click on the cDAQ and click on “reserve chassis”
  • D) Open changeVMchangeA.vi located in D:\Experiment\VIs\Wave Gauges Programs\SubVIs Insert IP Address (192.168.1.3 for smaller linear stage) and remote port (503) and run program
    These values are saved until the electrical boxes are turned off.
  • E) Move the linear stage (see the detailed instructions on Linear Stage Procedure) so that the middle (in vertical height) of the wave-gauges are positioned in the mean water level.
  • F) To set this point as 0, open LinearStageMaster.vi located in D:\Experiment\VIs\Linear Stage and insert IP Address of the same linear stage (192.168.1.3), and remote port (503), write “p 0” in the command window and run program.
  • Sanity check- write “pr p” in command window, run program. Check if the position equals 0 in data out.

Establishing gain and offset of the amplifier for each WG:

  • A) Turn on the wave-makers and create a corresponding amplitude value of the upcoming experiment
  • B) Open WG_ScopeSimple.vi located in D:\Experiment\Vis\Scope_Simple , insert 120 [Hz] and 1 [s] for the measurement time and run the program continuously
  • C) Adjust the offset and gain to obtain significant fluctuations of all WG channels in range of ±10 [V]
  • D) Shift all curves using offset to fluctuate around 0 [V]
  • E) Turn off the wave-makers

Calibration Process:

  • A) Open WaveGauges_Calibration.vi located in D:\Experiment\VIs\Wave Gauges Programs
  • B) Check if IP Address (192.168.1.3) and remote port (503) values of the linear stage are correct
  • C) Set the sampling rate and the measurement time (e.g. 1000 [Hz] & 10 [s]) for each calibration point
  • D) Set the calibration file name (.calib) and the file path to save the file
    (e.g.D:\Experiment\Flume measurements\CalibrationFiles)
  • E) Set the number of gauges to calibrate
  • F) Set the desired Z coordinates [mm] (absolute values) for the calibration (negative values ⇒ linear stage moves up; positive values ⇒ linear stage moves down).
    Increasing order of Z coordinates is preferred (e.g. -25 -20 -15 -10 -5 0 5 10 15 20 25)
    Double-check input values so that no damage is done to the stage, gauges and/or flume
  • G) Make sure the water is as still as possible in the flume and run program
  • Check calibration graphs for linear plots that matches the measured calibration points
  • Check calibration coefficients in *.calib file for at least 2 orders of magnitude difference between 2nd order coefficient (x2) and 1st order coefficient (x)

MANUAL WAVE GAUGE CALIBRATION PROCEDURE

For the Scottish wave gauges (unattached to the linear stage):

  • A) Set the middle (in vertical height) of the wave-gauges in the mean water level
  • B) Open ScottishWGScopeSimple.vi located in D:\Experiment\VIs\Scope_Simple
  • C) Set the sampling rate [Hz] and the measurement time [s] for each calibration measurement point
  • D) Make sure the water is as still as possible in the flume and run program
  • E) Write the value of voltage corresponding to this calibration point, as shown by the indicator
  • F) Move the wave-gauge up/down to the next calibration point position
  • G) Re-do steps D) to F) until enough calibration points are measured
  • H) Open Scottish_WG_Calib.m located in D:\Experiment\Flume measurements\Calibration_Files
  • I) Type in the values of calibration points voltage from lowest to highest value
  • J) Run the code

- Check that the number of rows equals the number of wave-gauges in the *.calib file
- Check calibration coefficients in *.calib file for at least 2 orders of magnitude difference between 2nd order coefficient (x2) and 1st order coefficient (x) in the last row

wave_gauges_calibration_procedure.txt · Last modified: by martin

Donate Powered by PHP Valid HTML5 Valid CSS Driven by DokuWiki