After wave gauges calibration process:
Turning on the equipment:
- Turn on the electrical boxes, Ethernet Switch, Amplifier & NI cDAQ-9184
- Turn on the DC power supply positioned on the instruments carriage and set to 5[V]
- Open changeVMchangeA.vi located in D:\Experiment\VIs\Wave Gauges Programs\SubVIs
- Insert IP Address (192.168.1.2 for larger linear stage) and remote port (503)
- run program
These values are saved until the electrical boxes are turned off
* To check if the DAQ Device is connected correctly:
open NI MAX ⇒ Devices and interfaces ⇒ Network Devices ⇒ cDAQ9184-1B560CB
right click on this cDAQ and select the “self-test“
- If completed successfully⇒ click on cards NI 9212, NI 9401, NI 9205 to enable them.
- If not successful⇒ right click on the cDAQ and click on ”reserve chassis”.
Finding and saving the safe position above the highest wave for the airflow sensors:
- Turn on the wind blower at frequency 40 [Hz]. Open ctrlSpeedBlower.vi located in D:\Experiment\VIs\Blower, insert 40 [Hz] and run the program
- Move the airflow sensors close to the water surface
- Open WaterSensorScopeSimple.vi located in D:\Experiment\VIs\Scope_Simple
This VI shows drop in voltage (~1.5-4[V]) when the sensor touches water
- Move the linear stage so that the exposed tips of the water sensor are in the water.
The drop of voltage will be evident on the graph/
Make sure the Pitot tube and static pressure probe are not touching the water
- Move the sensors up with a step of 3-4 [mm] and run the program for 30[sec]
- If there is a drop of voltage, repeat step e)
- If there is no drop of voltage, double check if you are in a safe position:
* run the program for 60 [sec] and if there is a drop of voltage, repeat step e)
* If there is no drop of voltage run the program for 120 [sec]
* If there is no drop of voltage proceed to next step, otherwise, repeat step e)
- Save this point as 0 by opening LinearStageMaster.vi located in \\D:\Experiment\VIs\Linear Stage and insert IP Address (192.168.1.2) and port (503).
Write “p 0” in the command window and run program
Sanity check
- Write “pr p” in command window, run program.
- Check if the position equals 0 in data out
- Open WindProfile.vi located in D:\Experiment\VIs\Wind Profile Program
Insert the following inputs:- Wind blower frequency:(e.g. 40, 32, 25, 18 [Hz])
- Seeking for zero water level:
* Sampling rate (e.g. 120 [Hz])
* Measurement time (e.g. 60 [s])
* Zero seek step (e.g. 3 [mm])
- Airflow sensors settings:
* Sampling rate (e.g. 1600 [Hz])
* Measurement time (e.g. 60 [s])
- Linear stage settings:
* remote port 503.
* IP Address 192.168.1.2
- Z coordinates (relative to 0[mm])
- Fetch (location of airflow sensors relative to drain holes [cm])
- Choose the calibration file
(located in D:\Experiment\Flume measurements\CalibrationFiles)
- Insert the Path to Data files
- Run the program.